Specification
|
Model |
SP-M1 |
|
Syringe Size |
10 ul – 1000 ul |
|
Power Requirement |
AC110V-220V,50Hz /60Hz orDC12V |
|
Pusher Advance Per Micro Step |
0.08 um |
|
Volume Per Micro Step |
132.4 nl with Unimetrics 1000ul syringe,
ID:4.61mm |
|
Minimum Flow Rate |
132.4 nl / min using Unimetrics 1000ul syringe, ID:4.61mm |
|
Maximum Flow Rate |
1.324 ml / min using Unimetrics 1000ul syringe, ID:4.61mm |
|
Linear Travel Rate |
Minimum: 7.94um/min |
|
Accuracy |
0.5%
(single infusion volume>30% total infusion volume ) |
|
Dimension |
Controller: 17.5 x 10.5 x 6.5 cm (LWH) Driver: 17.5 x 4.5 x 5 cm (LWH) |
|
User defined syringe ID |
0.01 mm-4.0 mm |
Flow Rate
|
Syringe (Unimetrics) |
Diameter |
Minimum |
Maximum |
|
10 ul |
0.46mm |
1.32 nl/min |
13.19 ul/min |
|
25 ul |
0.73mm |
3.32 nl/min |
33.2 ul/min |
|
50ul |
1.03mm |
6.61 nl/min |
66.1 ul/min |
|
100ul |
1.46mm |
13.28 nl/min |
132.8 ul/min |
|
250ul |
2.30mm |
32.97 nl/min |
329.7 ul/min |
|
500ul |
3.26mm |
66.25 nl/min |
662.5 ul/min |
|
1000ul |
4.61mm |
132.4 nl/min |
1324 ul/min |
|
|
|
Specification
|
Model |
SP-M2 |
|
Syringe Size |
1000 ul – 60ml |
|
Power Requirement |
AC110V-220V,50Hz /60Hz orDC12V |
|
Pusher Advance Per Micro Step |
0.08 um |
|
Volume Per Micro Step |
1Micro Step |
|
Minimum Flow Rate |
4.411ul / min using BectonDickinson 60ml syringe |
|
Maximum Flow Rate |
44.11 ml /
min using BectonDickinson 60ml syringe, ID:26.60mm |
|
Linear Travel Rate |
Minimum: 7.9um/min |
|
Accuracy |
0.5%
(single infusion volume>30% total infusion volume ) |
|
Dimension |
Controller: 17.5 x 10.5 x 6.5 cm (LWH) Driver: 24.5 x 10.0 x9. 5 cm (LWH) |
|
User defined syringe ID |
0.01 mm-40 mm |
Flow Rate
|
Syringe (Unimetrics) |
Diameter |
Maximum |
Minimum |
|
1ml |
4.7 |
1.377 |
137.7 |
|
3ml |
8.59 |
4.6 |
460.0 |
|
5ml |
11.99 |
8.96 |
896.2 |
|
10ml |
14.48 |
13.07 |
1307 |
|
20ml |
19.05 |
22.62 |
2262 |
|
30ml |
21.59 |
29.05 |
2905 |
|
60ml |
26.60 |
44.11 |
4411 |
|
|
Specification
|
Model |
SP-M3 |
|
Syringe Size |
1000 ul – 60ml |
|
Power Requirement |
AC110V-220V,50Hz /60Hz orDC12V |
|
Pusher Advance Per Micro Step |
0.08 um |
|
Volume Per Micro Step |
1Micro Step |
|
Minimum Flow Rate |
4.411ul / min using BectonDickinson 60ml syringe |
|
Maximum Flow Rate |
44.11 ml /
min using BectonDickinson 60ml syringe, ID:26.60mm |
|
Linear Travel Rate |
Minimum: 7.9um/min |
|
Accuracy |
0.5%
(single infusion volume>30% total infusion volume ) |
|
Dimension |
Controller: 17.5 x 10.5 x 6.5 cm (LWH) Driver: 24.5 x 10.0 x9. 5 cm (LWH) |
|
User defined syringe ID |
0.01 mm-40 mm |
Flow Rate
|
Syringe (Unimetrics) |
Diameter |
Maximum |
Minimum |
|
1ml |
4.7 |
1.377 |
137.7 |
|
3ml |
8.59 |
4.6 |
460.0 |
|
5ml |
11.99 |
8.96 |
896.2 |
|
10ml |
14.48 |
13.07 |
1307 |
|
20ml |
19.05 |
22.62 |
2262 |
|
30ml |
21.59 |
29.05 |
2905 |
|
60ml |
26.60 |
44.11 |
4411 |